Atlas Inspection

North America: 800-281-0650
Local/International: 206-774-8704
Contact Us by email>>

Customer Service Rating by LivePerson

Atlas Inspection Technologies

Enter your email address below for a free subscription to Inspection Innovations quarterly e-newsletter.



Basic Resolution Testing

A resolution test chart is used to allow quick and easy testing of the ability of an optical system to produce images with fine detail. The patterns are in groups which progressively get smaller. The smallest group for which the correct number of bars and spaces can be counted (in both directions) is the system resolution. Resolution is most commonly measured in cycles per mm (cy/mm), where a bar and a space equal one cycle. For example, a test image that has 5 cy/mm bars and spaces will have a bar 0.1mm wide and a space 0.1mm wide.

Bar type test charts are normally available on photo paper for reflection imaging with bars down to about 20 cy/mm, transparent flexible film to about 180 cy/mm, and chrome on glass or opal to 512 cy/mm.

When evaluating a system used for visual analysis, you need only to place a test pattern of known size in the field of view, then view it (by eye, camera, etc) to see the smallest group that is output. One of the oldest, most common charts is the USAF-1951 test chart which has some design features specified in Mil-Std-150A. It is available from Applied Image as the T-20 pattern and looks like:

t-20_pattern

One rather large drawback of this pattern is that the labels only indicate groups and elements, so a look up chart is needed to find out what the actual resolution is of an element.

The Applied Image T-21 solves this problem and still maintains compliance with Mil-Std-150A because the standard only specifies the density, number of bars, and progression of elements (not the labeling of arrangement).

t-21_pattern

The T-21 has labels which state the actual cy/mm of each element, so no look-up table is needed.

t-22_pattern

The T-22 further improves the pattern by arranging the elements in straight lines (much easier to locate, much easier to scan). Of course, it still meets the requirements of Mil-Std-150A. Finder squares have been added to help determine which element is being viewed, when the labels cannot be clearly seen.

Many other types of resolution test charts exist. The NBS-1010/ISO #2 (Applied Image T-10) is similar to the USAF-1951 chart, but has longer bars, and uses 5 bars instead of 3.

t-10_pattern

The T-10 is marked directly in cy/mm and is used extensively in the microfilm field.

Test charts can be made with a variety of contrast levels. Using a test chart with low contrast (a gray on white image is typical) is a more stringent test than high contrast because as lenses are introduced into an image path, one of the major degradations it causes is the reduction of contrast. Therefore, if the initial test object is lower contrast, any reduction in contrast will be a bigger part of the total signal.

Other types of test charts, such as the Sayce (T-60), are very well suited for scanning. Each line gets progressively smaller so a large range of line widths can be checked easily.

t-60_pattern

Sine patterns such as the M-5 array shown, allow a more sophisticated analysis. By scanning the sinusoidal density patches (of different frequencies) and evaluating them, the modulation transfer function (MTF) of a system can be determined.

array

Another way to evaluate optical systems is to start with a very sharply defined edge, with known characteristics (such as the dark gray on light gray slant edge of the QA-62), capture the image with a digital imager, then evaluate the image with the proper software. This also provides MTF data.

logchart

A complete catalog of standard test charts, scales, bar code standards, gray scale step tablets, and sine patterns can be found at www.aig-imaging.com. To contact Gary Reif at Applied Image call 585.482.0300 (ext 230) or email him at greif@appliedimage.com.